Trace VOC analyzer for real-time quantification with minimal user interaction
Detection limit: < 5 pptv
Mass Resolution: > 3500
Automated VOC Monitoring
The VOCentinel provides continuous, real-time quantitative results of a wide range of VOCs with only minimal user interaction required, making it an ideal choice for long-term air-quality, environmental and industrial monitoring applications.
Simple User Interface
The VOCentinel has a simple user interface for monitoring a predefined set of VOCs in realtime, including various relevant chemical classes, such as:
- amines
- aldehydes
- ketones
- alkenes
- monoterpenes
- siloxanes
- aromatics
Still, all gathered raw data is accessible for evaluation and further scientific research.
Robust Technology
Built on IONICON’s experience in robust industrial systems, the VOCentinel brings fully automated VOC monitoring to complex environmental applications. The instrument offers all benefits of a small footprint and low weight in a robust industry-proven setup. It is available as a benchtop instrument or as 19-inch rackmount version for easy integration into your monitoring station or mobile platforms.
Technical specifications
| Mass resolution | > m/Δm 3500 (FWHM) |
| High sample frequency | < 1min/sample point (combined data from multiple ionization modes) Full mass spectra in < 1 s |
| Detection limits for m/z 181 | < 5 pptv (in 60 s) |
| Linearity range | typically 10 pptv – 1ppmv |
| Weight | ~ 90 kg |
| Dimensions (w x h x d) | 48x80x66 cm
|
| Power supply | 115/230 V, typical operation: ~ 500 W |
Specifications are subject to change without prior notice.